General Information, Hematology

Test Update: Cell Count with Differential, Body Fluid

Beginning Tuesday, January 4, 2022, Spectrum Health Laboratories will include an automated neutrophil (PMN) count on Cell Count with Differential, Body Fluid (LAB210) orders for peritoneal body fluids.   This component will be displayed in Epic as an absolute PMN count, body fluid.

• The absolute PMN count in the peritoneal fluid is calculated by multiplying the total nucleated cell count by the percentage of PMNs in the differential.

• This calculation aids in the diagnosis of spontaneous bacterial peritonitis (SBP).

• The diagnosis of bacterial peritonitis is established by a positive peritoneal fluid bacterial culture and an elevated peritoneal fluid absolute PMN count (≥250 cells/uL).

• Epic will automatically calculate the absolute PMN count in peritoneal body fluids based on the manual differential data.

• This component ONLY calculates for Cell Count with Differentials (LAB210) on PERITONEAL BODY FLUIDS Continue Reading

Hemostasis/Coagulation

Update: PT/INR Resulting Change

The Spectrum Health Laboratory Coagulation Department announces the following changes occurring on 6.29.21 because of new instrumentation:

• New coagulation instrumentation will go-live on 6.29.21.

• This new instrumentation will align the Spectrum Health Regional Laboratory (in GR) with current Regional/Blodgett reportable upper limit for Protime/INR results.

• The current Protime (PT) upper limit of 120 seconds will change to 90 seconds. Any result that is greater than 90 seconds, will be reported as >90 seconds.

• The current INR upper limit of 12.0 will change to 8.0. Any result that is greater than 8.0 will be reported as >8.0.

Any questions can be directed to the Spectrum Health Regional Laboratory Coagulation Department using the Contact Us link above.

TEST INFORMATION 

PT/INR (Protime with INR): Epic Code #LAB320 | Interface code #10059 | CPT #85610