Compliance & Safety, Toxicology

Urgent Notice: Lead Testing

On May 17, 2017, the US Food and Drug Administration (FDA) issued a safety communication warning against using venous samples on Point of Care Magellan Diagnostics LeadCare Testing System devices (LeadCare, LeadCare II, LeadCare Plus, and LeadCare Ultra). Venous specimens appear to be at risk for having falsely low results. ¬†Only Capillary (“finger stick” or “heel stick”) specimen types should be used.

IF YOUR OFFICE IS USING ANY OF THE MAGELLAN LEADCARE TESTING SYSTEM DEVICES: please DISCONTINUE the use of VENOUS specimens for testing. Continue Reading